      美国TREK 520手持式静电电压测试仪
      应用: 人体或对象表面静电量测试,主要用于厂内静电产生情况的抽检
       

      Model 520 Series HAND-HELD Electrostatic Voltmeter

      The Model 520 Series includes the Model 520 and Model 523.
      The TREK Model 520 Hand-Held Electrostatic Voltmeter (ESVM) Series products provide accurate, noncontacting measurements of electrostatic surface voltage for ESD applications in either ionized or non-ionized environments. Industrial applications include measuring charge accumulation in the LCD manufacturing process or measuring static charge in semiconductor production. The Model 520 series ESVMs utilize a new measurement technique that overcomes the disadvantage of the typical hand-held fieldmeter by providing surface voltage measurements which are essentially independent of the sensor probe-to-measured surface spacing.

      Model:520-1 ESVM
      Over a spacing range of 5 mm to 25 mm, the 520 voltmeter provides better than a 5% voltage measurement accuracy, while the typical fieldmeter provides a voltage measurement error in excess of 50%. An optional voltage monitor output is available (Model:520-2). 

      Model 523 ESVM
      Over a spacing range of 30 mm to 60 mm, the 523 voltmeter provides greater than a 5% voltage measurement accuracy, while the typical fieldmeter provides a voltage measurement error in excess of 50%. 
      The 520 series provides a measurement system which complies with standardized calibration equipment traceable to the National Institute of Standards and Technology. The 520 series hand-held voltmeters employ a slender sensor probe design that provides flexibility to make measurements in many locations which are impossible to reach using the large, bulky sensing surfaces of the typical hand-held fieldmeter. The slender design also allows for higher spatial resolution capability by allowing the measurement of small surface areas. A "hold" pushbutton retains the displayed measured voltage value for recording purposes.
      ·  Model:520 Measurement Range 
      0 to ±2 kV DC.
      Model 523 Measurement Range 
      0 to ±20 kV DC.

      Model:520 Accuracy 
      ESVM Monitor (Model:520-2) 

      Better than ±5% of full scale over the entire recommended probe-to-surface separation range of 5 mm to 25 mm.
      Model 523 Accuracy 
      Better than ±5% of full scale over the entire recommended probe-to-surface separation range of 30 mm to 60 mm.

      Model:520 Probe-to-Surface Separation (recommended)
      5 mm to 25 mm.
      Model 523 Probe-to-Surface Separation (recommended)
      30 mm to 60 mm.
      Stability 
      Drift with Time 
      Less than 600 ppm/hour, noncumulative.
      Drift with Temperature 
      Less than 600 ppm/°C.